A cold spot compensation technique using a combination of trans-rectal ultrasonography and intraoperative computed tomography for interstitial permanent prostate brachytherapy: a single-arm prospective trial
J Contemp Brachytherapy 2018; 10, 1: 10–16
Online publish date: 2018/02/28
Get citation
Purpose
Trans-rectal ultrasonography (TRUS) is the standard imaging tool for interstitial prostate brachytherapy [1]. However, ultrasound is well known to be unsuitable for imaging implanted seeds, with computed tomography (CT) representing the standard for detecting seed positions and calculating post-implant dose-volume histograms (DVHs) [2,3].
The O-arm® surgical imaging system (Medtronic, Dublin, Ireland) was developed to provide real-time, intraoperative CT imaging with a large field-of-view. This system permits patients to remain in the lithotomy position even during image acquisition, because the bore diameter of this system (965 mm) is significantly larger than that of conventional CT systems (700-800 mm).
We combined TRUS with O-arm-based CT during surgery as a new strategy for intraoperative dosimetric evaluation. With this fusion dosimetry, we can compensate for the shortcomings of each modality, because accurate seed positions can be detected on CT, while accurate contours can be delineated by TRUS. Moreover, highly matched fusion images can be expected because of the removal of differences in patient and probe positions between the two modalities [4]. With this fusion dosimetry, we can check for the presence of cold spots and compensate with additional seeds during surgery.
The purpose of this prospective trial was to evaluate the efficacy of the cold spot compensation technique in interstitial prostate brachytherapy.
Material and methods
Patients
The institutional review board approved this single-arm prospective study (C15-911). Eligible participants were all adults 20 years old with localized prostate cancer without lymph node or distant metastases. Exclusion criteria were no indications for anesthesia, refusal to participate, or a need for additional external radiation therapy.
A total of 89 patients were assessed as eligible between November 2015 and July 2016. Twenty-four patients were excluded because they declined to participate (n = 4), did not meet the criteria (n = 13), had excess height or weight (n = 2), and other reasons (n = 5). As a result, 65 patients were treated using the cold spot compensation technique following TRUS-CT fusion. In addition, data from 100 consecutive patients treated by conventional techniques using only TRUS just before the start of this study (from October 2014 to November 2015) were collected as historical controls. Patient characteristics are shown in Table 1.
Seed implantation and image fusion technique
We have previously reported the details of our fusion technique for TRUS and O-arm-based CT [4]. Briefly, patients were placed in the lithotomy position inside the O-arm system. TRUS images of the whole prostate gland were acquired using a biplane transrectal ultrasound probe (HI VISION Preirus, Hitachi Aloka Medical, Tokyo, Japan). Treatment plans were developed on TRUS images using Variseed version 8.0.2 software (Varian Medical Systems, Palo Alto, CA). The prescribed dose to the prostate with a 3- to 5-mm margin was set as 145 Gy. Table 2 shows dosimetric parameters for the planning phase. Both loose-seed and intraoperatively built custom-linked (IBCL) seeds were used for patients in this study. Loose seeds were placed one by one transperineally through needles attached to a Mick applicator (Mick Radio-Nuclear Instrument, Mount Vernon, NY, USA). IBCL seeds were connected to each other using the quick-link system (CR BARD, Murray Hill, NJ, USA) and inserted through a relay system [5]. No dosimetric difference has been reported between loose seeds and IBCL seeds [6]. Two types of 125I source were used: either OncoSeed® model 6711 (GE Healthcare [Medi-Physics], Arlington Heights, IL, USA), or BrachySource® model STM125I (CR BARD, Murray Hill, NJ, USA). Source activities were 11.0 MBq or 13.1 MBq.
Following seed implantation, CT images were acquired using the O-arm system at 120 kV, 50 mA, and 200 mAs. Acquired CT images were transferred to the Variseed software. The end-fire probe was used as a landmark to fuse TRUS and O-arm-based CT images, due to clear recognition in both modalities. After image fusion, contours of the prostate, urethra, and rectal wall were copied from TRUS to CT images. The same contours were thus available on both modalities.
Additional seed implantation for cold spots
The software automatically detected seed position on TRUS-CT fusion images. The quality of detection, however, is not perfect, perhaps due to the low contrast resolution of O-arm-based CT images. Manual correction of automatically recognized seed positions was needed for all patients. Both a radiation oncologist and a urologist, reviewed dosimetry on TRUS-CT fusion images and checked for the presence of cold spots. If cold spots were found, additional seeds were implanted under TRUS monitoring according to planned positions on TRUS-CT fusion images (Figure 1). As our previous study suggested that a dose to 90% of prostate volume (D90) of 155 Gy on fusion images can be used as a surrogate for D90 > 170 Gy on 1-month follow-up CT analysis, we tried to reach around 155 Gy in the balance between urethral and rectal doses [4]. Parameters of D90 > 170 Gy (preferably > 180 Gy), V100 > 95%, V150 < 65%, and RV100 < 1 cc on 1-month CT were defined as the reference for “good quality implant” in this study, although there is no validated definition.
DVH analysis
DVHs were calculated from TRUS images, TRUS-CT fusion images, and 1-month CT images. Urethral contouring was based on the outer rim of the urethral catheter, except for 1-month CT, in which the center of the prostate was used as a surrogate for urethral position. The rectal wall including sphincter muscle was fully contoured on 1-month CT images, but only the anterior one-third excluding the lumen (body of the TRUS probe) was contoured on TRUS and TRUS-CT fusion. The urethra and rectum were contoured in the same slices as the prostate contour.
DVH parameters including D90, prostate volume receiving at least 100% dose (V100), prostate volume receiving at least 150% dose (V150), dose to 90% of urethral volume (UD90), dose to 30% of urethral volume (UD30), rectal volume receiving at least 100% dose (RV100), and rectal volume receiving at least 150% dose (RV150) were collected from TRUS, TRUS-CT fusion, and 1-month CT.
Urinary and rectal morbidity were assessed using the Radiation Therapy Oncology Group (RTOG) scale and the National Cancer Institute Common Terminology Criteria (NCI-CTC), version 4.
Statistical analysis
Statistical analyses were performed using R version 3.2.0 software (R Foundation, Vienna, Austria). Fisher’s exact test and independent t-test were used for comparisons of the examination arm and historical controls.
Results
After reviewing TRUS-CT fusion images, cold spots were detected in 32 of 65 patients (49%) and were subsequently compensated with additional seeds. Median number of additional seeds was 3 (range, 1-5). Table 3 shows changes in DVH parameters on TRUS-CT fusion images of the 32 patients before and after compensation.
When the examination arm was compared to historical controls, no significant difference in patient characteristics or planning parameters were seen between groups (Tables 1 and 2). Meanwhile, 1-month CT analysis revealed significant differences between the two groups in D90, V150, and UD30 (Table 4). Table 5 shows a comparison of percentages of patients with good quality implant between groups. The percentage of patients receiving a D90 over 180 Gy was significantly increased in the examination arm. Figure 2 shows a comparison of histograms for each DVH parameter between groups. Histograms basically showed a rightward-shift from the historical control group to the examination arm group. The percentage of patients receiving a D90 of 160-180 Gy was significantly reduced in the examination arm (19%) compared with the control arm (35%), although no significant differences in patients in other dose ranges were identified. Likewise, the percentage of patients receiving a UD30 of 175-200 Gy was significantly reduced in the examination arm (7%) compared with the control arm (28%).
However, mean operation time (from first needle insertion to end of operation) was significantly longer in the examination arm (64 min) than in historical controls (49 min, p < 0.001).
Table 6 shows acute toxicity until 12 months after treatment in the examination arm and historical controls. With a median follow-up of 18 months (range, 9-24 months) for the examination arm and of 19 months (range, 6-32 months) for historical controls, no significant difference was evident between groups except for urinary frequency. No instances of grade 3 or worse toxicity were encountered. Biochemical failure was seen in no patients from the examination arm and in 4 historical controls.
Discussion
By compensating for cold spots during surgery, a significant percentage of patients were salvaged from poor dosimetry on 1-month CT analysis. Our system enables checking for the presence of cold spots during surgery and permits implantation of additional seeds without second anesthesia. We believe that this system is useful not only for experts, but also for beginners lacking experience in brachytherapy, providing an extra margin of safety for favorable dosimetry.
Adaptive planning for interstitial permanent prostate brachytherapy using intraoperative CT was already reported from the Netherlands in 2007 [7]. The same group recently reported not only improvement of dosimetric results [8], but also of biochemical control with their technique from a large database of over 1,600 patients [9]. Their surprising improvement in clinical results (28% increase in 7-year biochemical control for high-risk patients) encouraged us to continue with this technique and to promote its spread to other hospitals.
As shown in Figure 2, however, patients with a low D90 did not disappear with our technique. One of the shortcomings of our technique is the lack of a method for adapting to changes in prostate volume. Although prostate volume usually decreases from the intraoperative phase to 1-month CT, some patients inversely show an increase, or no change compared to the intraoperative phase. These patients inevitably show a low D90 on 1-month CT because we set dosimetric parameters in the planning phase on the assumption that prostate volume would decrease by 1 month after treatment [4]. In addition, prostate contours might be ambiguous in some patients and contouring errors might have some effects on 1-month CT analysis [4].
The other key shortcoming of our technique was operation time. Our software takes a long time to recognize implanted seed positions from TRUS-CT fusion images. Furthermore, the quality of recognition is imperfect and manual correction is needed for all patients. Operation time was thus extended by 15 min compared to historical controls. However, we believe this represents an acceptable cost, considering the utility of cold spot compensation.
One of the strong points of our technique is highly matched fusion of TRUS and CT images, due to the identical patient and probe positions. In reports on the utility of intraoperative CT from other institutions [8,9,10,11], positions of patients and probes changed from operation to intraoperative CT acquisition. This can result in significant changes to prostate shape [12,13]. The positions of the legs also seem certain to have some effect on prostate shape. Although these differences may seem small, such small differences might cause large differences in the results of brachytherapy because of the very sharp dose gradient. The bore size of the O-arm system may be problematic for American or European individuals, who are relatively taller than Japanese population. Although most of our patients could remain in the lithotomy position, two patients taller than 190 cm or weighing over 137 kg could not undergo O-arm CT due to positioning difficulties.
Toxicity was acceptable with our compensation technique. The crude rate of urinary frequency was lower in the examination arm than in historical controls. Although the reasons for this were not clear, the short follow-up of the examination arm might have underestimated urinary frequency.
Conclusions
This prospective study showed that our cold spot compensation technique can salvage a significant percentage of patients from poor dosimetry with acceptable toxicity and operation time.
Disclosure
Authors declare no conflict of interest. Dr. Ishiyama reports personal fees from Medicon, Inc., personal fees from Nihon Medi-Physics Co., Ltd., outside the submitted work.
References
1. Polo A, Salembier C, Venselaar J et al. Review of intraoperative imaging and planning techniques in permanent seed prostate brachytherapy. Radiother Oncol 2010; 94: 12-23.
2. Nag S, Bice W, DeWyngaert K et al. The American Brachytherapy Society recommendations for permanent prostate brachytherapy postimplant dosimetric analysis. Int J Radiat Oncol Biol Phys 2000; 46: 221-230.
3. Davis BJ, Horwitz EM, Lee WR et al. American Brachytherapy Society consensus guidelines for transrectal ultrasound-guided permanent prostate brachytherapy. Brachytherapy 2012; 11: 6-19.
4. Ishiyama H, Sekiguchi A, Satoh T et al. Dosimetry of permanent interstitial prostate brachytherapy for an interoperative procedure, using O-arm based CT and TRUS. J Contemp Brachytherapy 2016; 8: 7-16.
5. Zauls AJ, Ashenafi MS, Onicescu G et al. Comparison of intraoperatively built custom linked seeds versus loose seed gun applicator technique using real-time intraoperative planning for permanent prostate brachytherapy. Int J Radiat Oncol Biol Phys 2011; 81: 1010-1016.
6. Ishiyama H, Satoh T, Kawakami S et al. A prospective quasi-randomized comparison of intraoperatively built custom-linked seeds versus loose seeds for prostate brachytherapy. Int J Radiat Oncol Biol Phys 2014; 90: 134-139.
7. Westendorp H, Hoekstra CJ, van’t Riet A et al. Intraoperative adaptive brachytherapy of iodine-125 prostate implants guided by C-arm cone-beam computed tomography-based dosimetry. Brachytherapy 2007; 6: 231-237.
8. Westendorp H, Hoekstra CJ, Immerzeel JJ et al. Cone-beam CT-based adaptive planning improves permanent prostate brachytherapy dosimetry: An analysis of 1,266 patients. Med Phys 2017; 44: 1257-1267.
9. Peters M, Smit Duijzentkunst DA, Westendorp H et al. Adaptive cone-beam CT planning improves long-term biochemical disease-free survival for 125 I prostate brachytherapy. Brachytherapy 2017; 16: 282-290.
10. Zelefsky MJ, Worman M, Cohen GN et al. Real-time intraoperative computed tomography assessment of quality of permanent interstitial seed implantation for prostate cancer. Urology 2010; 76: 1138-1142.
11. Kaplan ID, Meskell P, Oldenburg NE et al. Real-time computed tomography dosimetry during ultrasound-guided brachytherapy for prostate cancer. Brachytherapy 2006; 5: 147-151.
12. Ishiyama H, Kitano M, Satoh T et al. Difference in rectal dosimetry between pre-plan and post-implant analysis in transperineal interstitial brachytherapy for prostate cancer. Radiother Oncol 2006; 78: 194-198.
13. Seppenwoolde Y, Kolkman-Deurloo IK, Sipkema D et al. HDR prostate monotherapy: dosimetric effects of implant deformation due to posture change between TRUS- and CT-imaging. Radiother Oncol 2008; 86: 114-119.
Copyright: © 2018 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License ( http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
|
|