Day to day treatment variations of accelerated partial breast brachytherapy using a multi-lumen balloon

Hsiang-Chi Kuo
,
Keyur J. Mehta
,
Linda Hong
,
Ravindra Yaparpalvi
,
Leslie L. Montgomery
,
William Bodner
,
Wolfgang A. Tomé
,
Shalom Kalnicki

J Contemp Brachytherapy 2014; 6, 1: 68–75
Online publish date: 2014/04/03
View full text Get citation
 
Purpose: To evaluate the variations of multi-lumen balloon (MLB)-based brachytherapy from simulation day to treatment day and their dosimetric impacts during accelerated partial breast irradiation (APBI).

Material and methods: A total of 42 CT images scanned from seven patients were evaluated with regards to daily variation due to of: 1) internal uncertainty: size and shape of balloon, seroma volume; 2) geometrical uncertainty-random: length of each catheter was measured for each fraction (total 70); 3) geometrical uncertainty-systematic: virtual systematic errors were tested by offsetting dwell positions. The original plans (as group A) had a mean value of 96.8% on v95 of the PTV_Eval. Plans were rerun (as group B) such that the mean value of the v95 was relaxed to 90.4%. By applying the reference plan to each daily CT image, variations of target coverage under different sources of error were evaluated.

Results: Shape and size of the balloon had means of < 1 mm decreased in diameter and < 0.4 cm3 decreased in volume; the mean seroma volume increased by 0.2 cm3. This internal variation has a mean of < 1% difference for both V90 and v95. The geometrical uncertainty made a mean deviation of 2.7 mm per root of sum of square. It caused the degradations of V90 and v95 by mean values of 1.0% and 1.2%, respectively. A systematic error of 3 mm and 4 mm would degrade both of V90 and v95 by 4% and 6%, respectively. The degradations on target coverage of the plans in group A were statistically the same as those in group B.

Conclusions: Overall, APBI treatments with MLB based brachytherapy are precise from day to day. However, minor variation due to daily treatment uncertainties can still degrade tumor bed coverage to an unacceptable coverage when v95 of the original plan is close to 90%.
keywords:

APBI, brachytherapy, breast cancer, Contura, multi-lumen balloon, treatment uncertainty

 