Original paper
Dosimetric impacts of applicator displacements and applicator reconstruction-uncertainties on 3D image-guided brachytherapy for cervical cancer

Joshua Schindel
,
Winson Zhang
,
Sudershan K. Bhatia
,
Wenqing Sun
,
Yusung Kim

J Contemp Brachytherapy 2013; 5, 4: 250-257
Online publish date: 2013/12/18
Article file
- dosimetric.pdf  [0.37 MB]
Get citation
 
 

Purpose

3D image-based treatment planning for gynaecological (GYN) cancer is a recent innovation [1-7] where magnetic resonance imaging (MRI) and computerized tomography (CT) are used to accurately define target volumes, organs at risk (OARs), and applicator position [5]. Dose calculations are dependent on the geometrical accuracy of the source position relative to target volumes and OARs [8]. The challenge with 3D image guidance in most centers is that the patient must be transferred out of the high-dose-rate (HDR) brachytherapy (BT) operating room for imaging. The patient transfer between applicator implantation, imaging, and radiation delivery can cause applicator displacements. These applicator displacements lead to dosimetric errors to the target volume and OARs [4,8]. Gerzten et al. have researched applicator shifts due to patient movement from simulation to HDR treatment rooms [9]. However, applicator displacements were evaluated on orthogonal radiographs, and no dosimetric impact on target volumes and OARs were measured. Real time applicator monitoring systems using Microsoft’s KINECTTM system (Microsoft Corp. Redmond, WA, USA) was recently introduced as a possible solution to monitor intracavitary applicator displacement with less than or equal to 1 mm resolution for applicator displacement during additional 3D scans in a different room [10]. However, to set applicator displacement limits, it is important to first understand its dosimetric impact. No study has yet presented the dosimetric impact of intracavitary applicator displacements for both conventional, point A plans and 3D image-guided, conformal plans.

Another major challenge facing 3D image guided planning is to reconstruct the source-pathway on the 3D image, especially those on MRI. Exclusive MRI-based applicator reconstruction methodologies were introduced for different applicators for cervical cancer such as plastic tandem-and-ovoids (T&O) [11,12], plastic tandem-and-ring (T&R) applicators [13], titanium T&R applicators [13], titanium Rotterdam applicators [14], and Vienna applicators (plastic T&R plus supplementary needles) [3]. Among the first applicator reconstruction accuracy studies [4,6,11-15] on MRI, two reports [7,14] presented absolute reconstruction uncertainties using phantom studies for plastic and titanium T&R [7] and titanium Rotterdam applicators [14], while other reports presented relative comparison data using different reconstruction methods [15] or different MRI scanning protocols [12]. When a plastic applicator is used for MRI guided BT, the signals of the MRI-marker catheters [11] are able to guide clinicians to reconstruct applicator position based on their hyposignals. The use of titanium applicators using MRI makes their reconstruction much more difficult due to the artifacts they create. Recently a novel Marker-Flange [16] was introduced to improve the reconstruction accuracy of titanium applicators when used with applicator library. GEC-ESTRO (Groupe Européen de Curiethérapie-European Society for Therapeutic Radiology and Oncology) reported the considerations and pitfalls in new applicator commissioning, and the reconstruction uncertainties this entailed with a 3D image [4]. However, 3D imaging reconstruction uncertainties should always be validated in each clinic when a clinic implements a 3D image guided BT program, since reconstruction uncertainties are associated with each institutional scanning protocols, the type of CT or MRI scanner, the type of applicator, applicator material, treatment planning system (TPS), and planning workflow. For instance, when a titanium applicator is used with a CT scanner, the titanium applicator on the CT image can have a bigger diameter (e.g. 6 mm versus 3.2 mm) due to the Hounsfield unit changes between the titanium and adjacent tissue. Namely, when a default-window level is used, if a clinician defines the end of bright white signal on CT as a tip of tandem, this can cause systematic reconstruction errors that result in systematic dosimetric errors. It is important to understand the estimated dosimetric impact of the reconstruction uncertainties. For MRI-guided, conformal BT plans, Tanderup et al. [8] presented the dosimetric impact of T&R on target volumes and OARs through by virtually shifting or rotating each tandem and ring before performing dose calculations of volume optimization. No study has yet presented the dosimetric impact of reconstruction accuracy for conventional, point A plans. In this study, we present the dosimetric impact of applicator displacement and applicator reconstruction-uncertainties on target volumes and OARs for both conventional, point-A plans, and 3D image guided conformal plans.

Material and methods

HDR brachytherapy planning: point A and MRIG-CBT plans



Conventional point A plans were clinically generated on for MRI, following the recommendations of the American Brachytherapy Society consensus guidelines [17,18]. The BrachyVisionTM (Varian Medical Systems Inc., Palo Alto, CA, USA) TPS was used. A high resolution (3.0 Tesla) Siemens MAGNETOM Trio 3T MR scanner (Siemens Medical Solutions Inc., Erlangen, Germany) was used for all MR scanning following the same GYN MRI protocols as previously reported [19]. Twenty randomly selected HDR cases were retrospectively studied after Institutional Review Board approval was secured. A titanium Fletcher-Suit-Delclos T&O applicator (Varian Medical Systems, Inc.) was used for all implants, along with gauze packing. The position of the bladder foley at the most inferior bladder neck was checked after foley insertion, and also re-checked on MRI. Orthogonal radiographs using a C-arm were taken before and after each MRI scan (Pre- and Post-X-ray). The institutional protocols for external beam radiotherapy and HDR BT fractionation schemes were detailed in previous studies [7,20-22]. High-risk clinical target volumes (HR-CTV), of the rectum, bladder, and sigmoid were retrospectively delineated as per GEC-ESTRO recommendations [2,3]. For each point A plan, a MRI-guided, conformal BT (MRIG-CBT) plan was retrospectively generated using the hybrid-inverse optimization detailed in a previous study [7].



Simulating T&O displacements and T&O reconstruction-uncertainties



To simulate T&O displacements, in unshifted, original point A plans (Figs. 1B,D) and MRIG-CBT (Figs. 1C,E), a whole T&O (Fig. 1A) was virtually shifted in a cranial (+) or caudal (–) direction in ± 1.5, ± 3, ± 5, ± 6, ± 7.5, ± 10, and ± 20 mm increments after dose calculations (Fig. 1). Simply put, the whole T&O was shifted cranio-caudally while dwell times and positions were kept unchanged (the unshifted plan) in order to model the physical T&O displacement that can occur during patient transfer between the BT treatment room and CT or MRI scanning room. Given that, the T&O is tightly packed with gauze and tandem angle limits torque, the T&O rarely rotates, but is often displaced in a cranial-caudal direction during 3D scanning. For the purpose of this study, only the T&O displacement in cranial-caudal direction was simulated in this study. Absolute dosimetric changes due to virtual shifting were measured for both point-A and MRIG-CBT plans between an unshifted and a shifted plan.

Regarding T&O reconstruction-uncertainty simulations, a tandem and two ovoids were shifted along their central axis (see Fig. 2A), since reconstruction uncertainties exist mainly along their central axis when an applicator library is used. Unlike that the ring reconstruction uncertainties for T&R that occur when it rotates, ovoid reconstruction uncertainties for T&O occur in a posterior-anterior direction along the central axis of the ovoid-tandem (Fig. 2A). We defined (+) direction and (–) direction for either tandem or ovoid motion when a source-wire travels longer (shorter), along its source-pathway. As a result, a tandem was virtually shifted in a cranial (+) or caudal (–) direction, while each ovoid was virtually shifted in a posterior (+) or anterior (–) direction. For the purpose of this study, a tandem and both ovoids were virtually shifted in increments of ± 1.5, ± 3.0, ± 4.5, ± 6.0, ± 7.5, and ± 10 mm. For instance, a –10 mm shifted plan represents a tandem is shifted 10 mm in a caudal direction and each ovoid is shifted 10 mm in an anterior direction as depicted in Fig. 2A. After virtual shifting was completed, dose calculations of either a point A plan or a MRIG-CBT plan was performed, given applicator reconstruction. For the point A plan (Fig. 2B), reference optimization lines were created based upon the shifted applicator and optimization was done using those regenerated reference lines. For the MRIG-CBT plan (Fig. 2C), a hybrid-inverse optimization [7,23] was performed after each shift. It is noted that the isodose lines (i.e. dwell times) are delivered at the correct dwell positions, even when an applicator is reconstructed inaccurately (i.e. a shift of a couple of mm from the true position). To simulate delivered-isodose lines, we shifted the calculated isodose lines and the T&O back to its original position (Figs. 2 D,E) which we then used as a reference plan. Each shifted plan has its own reference plan. Afterward, we compared the dose parameters between the shifted plans (Figs. 2 B,C) and their reference plans (Figs. 2 D,E).

Dosimetric changes on dose-volume parameters of the HR-CTV (D100 and D90), the rectum (D2cc), the bladder (D2cc), and the sigmoid (D2cc) were recorded for both point A and MRIG-CBT plans. In addition, dosimetric changes of point A (left and right points) and ICRU report #38 defined, rectal and bladder points were quantified. To normalize the prescription dose differences, the percent dose differences were calculated for all dosimetric metrics. The p-values were calculated using a two-tailed t-test.

Results

Dosimetric impact of T&O displacements



The dosimetric impact of simulated T&O displacements on sigmoid (p < 0.0005), bladder (p < 0.0001), HR-CTV (p < 0.0036), and point A (p < 0.0015) were significantly larger for the MRIG-CBT plans than the point A plans. MRIG-CBT plans data in Table 1 clearly presented more dark (dosimetric changes ≥ 15%) and gray (dosimetric changes ≥ 10%) regions than point A plans as seen in Table 2. Rectal D2cc values were observed as the most sensitive to change due to the T&O displacement across all dosimetric metrics, regardless of point A (p < 0.013) or MRIG-CBT plans (p < 0.0277). Even a 3 mm T&O displacement caused an average 10-12% dosimetric change on rectal D2cc for both point A plans (Table 2) and MRIG-CBT plans (Table 1). A 5 mm T&O displacement resulted in clinically unacceptable dosimetric changes on rectal D2cc for point A plans (average 17-21%, Table 2) and MRIG-CBT plans (average 16-22%, Table 1). A 1.5 mm T&O displacement solely presented less than a 10% dosimetric change (average 6%) for both point A (Table 2) and MRIG-CBT plans (Table 1). The D2cc value for rectum showed a larger dosimetric change than the ICRU rectal point doses for point A (p < 0.0389) and MRIG-CBT plans (p < 0.0699). A more than 5 mm T&O displacement in a cranial direction resulted in a greater than 10% change in bladder D2cc for MRIG-CBT plans. Bladder (D2cc) of point A plans presented a relatively low dosimetric change (less than 10%) up to 7.5 mm T&O displacement. Sigmoid (D2cc) also had a relatively low dosimetric change (less than 10%) up to 10 mm and 7.5 mm T&O displacement for both point A and MRIG-CBT plans, respectively. HR-CTV (D100 and D90) was also found to change less than 10% for up to 10 mm (7.5 mm) of the T&O displacement for MRIG-CBT plans (point A plans). The point A doses were found to be less than 10% of the changes for up to 10 mm in the T&O displacement for point A plans, since point A locates on the reference optimization lines. However, more changes in the point A doses were observed (11-17% changes per 10 mm displacement) in MRIG-CBT plans, since they don’t have typical pear-shape isodose lines, but have conformal isodose lines.



Dosimetric impact of T&O reconstruction-uncertainties



Our findings suggest that in order to avoid more than 10% dosimetric changes in any dosimetric parameter either point A (Table 4) or MRIG-CBT plans (Table 3), had to have less than or equal to ± 3.0 mm reconstruction-uncertainty. For conventional point A plans, the rectum (D2cc, ICRU point) and bladder (ICRU point) presented greater dosimetric changes due to the reconstruction uncertainties (Table 4). The rectal D2cc value recorded was the most sensitive to the T&O reconstruction. To ensure that less than 10% dosimetric change occurred in rectal D2cc a value of less than or equal to ± 1.5 mm in reconstruction uncertainty was required. A 3 mm reconstruction uncertainty in the cranial direction caused an average 15% dosimetric change in rectal D2cc. ICRU bladder point presented a greater than 10% change due to the 4.5 mm T&O reconstruction uncertainty in the caudal direction. ICRU bladder point doses were observed to be more sensitive than bladder D2cc due to the close proximity of the ovoids.

The impact of T&O reconstruction-uncertainties was the most at ICRU bladder point doses (Table 3). A 4.5 mm reconstruction-uncertainty caused 10% of dosimetric changes to ICRU bladder point doses. A 7.5 mm reconstruction-uncertainty in the caudal direction resulted

in an average 10% change in ICRU rectal point doses. HR-CTV and point A doses presented relatively small changes, less than 10% per ± 7.5 mm T&O reconstruction uncertainties, for either point A plans or MRIG-CBT.

Discussion

To the best of our knowledge, this is the first study of dosimetric impact due to T&O applicator displacements for both point A and MRIG-CBT plans. In addition, this is the first study presenting reconstruction uncertainty impact for both point A plans and MRIG-CBT plans. Regarding the impact of T&O displacements, MRIG-CBT plans demonstrate significantly larger dosimetric changes on sigmoid (p < 0.0005), bladder (p < 0.0001), HR-CTV (p < 0.0036) than those of point A plans. It is mainly due to conformal isodose line target volumes and OARs. However, regardless of MRIG-CBT plans and point A plans, rectal D2cc changed more than 10% for ± 3 mm T&O displacements. Our findings agree with those of Tanderup et al. [8]. They found a mean change of 15% for the D2cc values of the rectum and bladder due to 3 mm cranial-caudal displacement, when T&R or combined T&R + needles applicators were used for MRIG-CBT plans. Based upon our findings, a ± 1.5 mm T&O displacement allowed all dosimetric metrics to change less than 10%. For the purpose of this study, we simulated motion in both cranial and caudal directions, but in clinical practice, T&O displacement in the caudal direction typically occurs due to movement of the patient anatomy. As yet, no study presents intracavitary applicator displacements due to patient transfer for 3D image scans and their dosimetric impacts on dose-volumetric parameters using 3D images. Gerszten et al. reported an avg. 5 mm (range, 3 to 12 mm) displacement when using an immobilization device (see Fig. 2 in [9]) and 2D radiograph images. It is important to minimize intracavitary applicator displacement when transferring patients. One way to limit patient transfer for 3D image acquisition for image-guided HDR would be to install a CT scanner in the HDR BT room. However, CT provides limited information about macroscopic tumor volume for cervical cancer that MRI is able to provide [5]. Another possible solution is to utilize one of the commercial immobilization systems such as those available from Zephyr patient transport system (DIACOR Inc., Salt Lake City, Utah, USA) using air-bearing techniques or MedTrak (MedTrak, LLC, Madison, Wisconsin, USA) that uses ceiling rails. Those systems are expected to reduce applicator displacement due to patient transfer. However, none of these support devices address applicator immobilization with respect to the patient. Our T&O support device was securely attached to the patient, allowing it to move as the patient does. In contrast to this, in some immobilization systems the immobilization device (see Fig. 2 in [9]) is secured to a transfer-board or a table, which can cause damage like uterine perforation due to even small patient movements. Applicator displacement can occur due to one or many reasons including anatomy changes due to insufficient anesthesia, suboptimal vaginal packing, and vibrations from the transfer table while in motion between rooms. Thus, the use of an applicator immobilizing device with respect to a patient is recommended even when a commercial immobilization system is in place. Even when using immobilization systems, applicator displacement must be carefully monitored when a patient is moved to different rooms for MRI or CT imaging. Studies for a robust immobilization system that limits motion, yet is still cost effective are expected to proliferate as 3D image-guided BT is more widely practiced in clinics.

Recently, planning uncertainties on 3D image guided BT have been widely explored in the literature. An sample of some of the reports include: the impact of uncertainties on planning and prescription [25], inter- and intra-fractional anatomical variations [26] and their dosimetric impact [27], especially for pulse-dose-rate (PDR) BT, and critical structure movement between planning MRI and secondary MRI before HDR [28] treatment. The overall D2cc value of multiple BT fractions is obtained under the assumption that the approximate D2cc location remains the same through multiple fractions. Jamema et al. [29] evaluated inter-application variation of doses and spatial location of OAR D2cc, and found that the D2cc locations were quite stable for bladder and rectum. However, significant topographical changes were reported for the sigmoid [29]. For the bladder, the current simple DVH parameter addition was also validated as compared to the overall DVH through deformable image registration (DIR) by Andersen et al. [30]. They found a current D2cc addition for the bladder provides a good estimate, whereas D0.1cc is less robust when compared to the values obtained using DIR. In relation to target volume delineation uncertainties, Petrič et al. found 0.76 (76% agreement) and 0.72 for HR-CTV for its simultaneous truth and performance level estimation, and expert consensus, respectively [31]. HR-CTV was found to be more robust than gross target volume and intermediate risk CTV. The dosimetric impact of delineation uncertainty for HR-CTV and OARs were quantified by Hellebust et al. [32]. They found that interobserver delineation variability caused ± 5 Gya/b = 10 of equivalent dose in 2 Gy fraction (EQD2) for HR-CTV D90 and ±2-3 of EQD2 for OAR D2cc. Treatment planning technique variations among four T&O user institutions were compared for point A plans, T&O MRIG-CBT plans, and T&O + needles MRIG-CBT plans [33]. Results indicated that adding interstitial needles to MRIG-CBT plans resulted in higher tumor coverage, but increased the variety in dose distributions between the institutions [33]. The uncertainties of performing two HDR fractions within one applicator implant were reported by Lang et al. [34]. They concluded that geometric differences between applicator, target, and OAR resulted in an overall dosimetric change (± 1.2 Gya/b = 10 EQD2 of HR-CTV D90 and ± 0.7 – 1.1 Gya/b = 3 EQD2 for OAR D2cc) that could be of minor relevance in regard to clinical dose volume constraints.

To account for geometric uncertainties of applicator reconstruction or displacements, Tanderup et al. [8,24] proposed a margin in the cranial direction only along a tandem (see Fig. 1 in [24]). Current 3D-imaging guidelines for GYN cancer [2-4,18] do not have margins to account for the uncertainties due to applicator displacements and/or an applicator reconstruction inaccuracy [4,8]. As highlighted by Tanderup et al. [24], the general planning target volume (PTV) concept should not be used to compensate for the uncertainties in 3D image guided brachytherapy (see Fig. 2 in [24]). They recommended not creating a general PTV margin, especially in the lateral and anterior-posterior directions as that could lead to a significant overall dose escalation: ~8%/mm within the margin applied. Dose prescription or normalization to PTV was also strongly discouraged in the study [24].

Conclusions

Regarding the dosimetric impact of T&O displacement, rectal D2cc values were observed as being most sensitive to change due to the T&O displacement among all dosimetric metrics regardless of point A (p < 0.013) or MRIG-CBT plans (p < 0.0277). To avoid greater than 10% changes, no more than ±1.5 mm T&O displacements were allowed for both point A and MRIG-CBT plans. The dosimetric impact of simulated T&O displacements on sigmoid (p < 0.0005), bladder (p < 0.0001), HR-CTV (p < 0.0036), and point A (p < 0.0015) were significantly larger in MRIG-CBT plans than point A plans. Regarding the dosimetric impact of T&O reconstruction-uncertainties, less than a ± 1.5 mm reconstruction-uncertainties was required in order to avoid a greater than 10% dosimetric change to point A plans.

Acknowledgements

This work was partly supported by a research collaboration grant from Varian Medical System, Inc.

References

 1. Viswanathan AN, Erickson BA. Three-dimensional imaging in gynecologic brachytherapy: a survey of the American Brachytherapy Society. Int J Radiat Oncol Biol Phys 2010; 76: 104-109.

 2. Haie-Meder C, Pötter R, Van Limbergen E et al. Recommendations from gynaecological (GYN) GEC-ESTRO Working Group (I): concepts and terms in 3D image based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV. Radiother Oncol 2005; 74: 235-245.

 3. Pötter R, Haie-Meder C, Van Limbergen E et al. Recommendations from gynaecological (GYN) GEC ESTRO working group (II): concepts and terms in 3D image-based treatment planning in cervix cancer brachytherapy-3D dose volume parameters and aspects of 3D image-based anatomy, radiation physics, radiobiology. Radiother Oncol 2006; 78: 67-77.

 4. Hellebust TP, Kirisits C, Berger D et al. Recommendations from gynaecological (GYN) GEC-ESTRO working group: considerations and pitfalls in commissioning and applicator reconstruction in 3D image-based treatment planning of cervix cancer brachytherapy. Radiother Oncol 2010; 96: 153-160.

 5. Viswanathan AN, Dimopoulos J, Kirisits C et al. Computed tomography versus magnetic resonance imaging-based contouring in cervical cancer brachytherapy: results of a prospective trial and preliminary guidelines for standardized contours. Int J Radiat Oncol Biol Phys 2007; 68: 491-498.

 6. Anderson J, Huang Y, Kim Y. Dosimetric impact of point A definition on high-dose-rate brachytherapy for cervical cancer: evaluations on conventional point A and MRI-guided, conformal plans. J Contemp Brachytherapy 2012; 4: 241-246.

 7. Anderson J, Xia J, Flynn RT et al. High resolution (3 Tesla) MRI-guided conformal brachytherapy for cervical cancer: consequences of different high-risk CTV sizes. J Contemp Brachytherapy 2013; 5: 101-109.

 8. Tanderup K, Hellebust TP, Lang S et al. Consequences of random and systematic reconstruction uncertainties in 3D image based brachytherapy in cervical cancer. Radiother Oncol 2008; 89: 156-163.

 9. Gerszten K, Paul C, Kin G et al. High dose rate tandem and ring applicator movement with patient transfer from simulation to treatment room. J Brachytherapy Int 1998; 14: 15-20.

10. Xia J, Waldron T, Kim YA. Real-Time Applicator Position Monitoring System (RAPS) for High-Dose-Rate Intracavitary Brachytherapy. Med Phys 2013; 40: 465.

11. Perez-Calatayud J, Kuipers F, Ballester F et al. Exclusive MRI-based tandem and colpostats reconstruction in gynaecological brachytherapy treatment planning. Radiother Oncol 2009; 91: 181-186.

12. De Leeuw AA, Moerland MA, Nomden C et al. Applicator reconstruction and applicator shifts in 3D MR-based PDR brachytherapy of cervical cancer. Radiother Oncol 2009; 93: 341-346.

13. Haack S, Nielsen SK, Lindegaard JC et al. Applicator reconstruction in MRI 3D image-based dose planning of brachytherapy for cervical cancer. Radiother Oncol 2009; 91: 187-193.

14. Petit S, Wielopolski P, Rijnsdorp R et al. MR guided applicator reconstruction for brachytherapy of cervical cancer using the novel titanium Rotterdam applicator. Radiother Oncol 2013; 107: 88-92.

15. Berger D, Dimopoulos J, Pötter R et al. Direct reconstruction of the Vienna applicator on MR images. Radiother Oncol 2009; 93: 347-351.

16. Schindel J, Muruganandham M, Pigge FC et al. Magnetic resonance imaging (MRI) markers for MRI-guided high-dose-rate brachytherapy: novel marker-flange for cervical cancer and marker catheters for prostate cancer. Int J Radiat Oncol Biol Phys 2013; 86: 387-393.

17. Viswanathan AN, Thomadsen B and Committee ABS. CCR American Brachytherapy Society consensus guidelines for locally advanced carcinoma of the cervix. Part I: general principles. Brachytherapy 2012; 11: 33-46.

18. Viswanathan AN, Beriwal S, De Los Santos JF et al. American Brachytherapy Society consensus guidelines for locally advanced carcinoma of the cervix. Part II: high-dose-rate brachytherapy. Brachytherapy 2012; 11: 47-52.

19. Kim Y, Muruganandham M, Modrick JM et al. Evaluation of artifacts and distortions of titanium applicators on 3.0-Tesla MRI: feasibility of titanium applicators in MRI-guided brachytherapy for gynecological cancer. Int J Radiat Oncol Biol Phys 2011; 80: 947-955.

20. Anderson J, Huang Y, Kim Y. Dosimetric impact of point A definition on high-dose-rate brachytherapy for cervical cancer: evaluations on conventional point A and MRI-guided, conformal plans. J Contemp Brachytherapy 2012; 4: 241-246.

21. Rockey WM, Bhatia SK, Jacobson GM et al. The dosimetric impact of vaginal balloon-packing on intracavitary high-dose-rate brachytherapy for gynecological cancer. J Contemp Brachytherapy 2013; 5: 17-22.

22. Sun W, Bhatia SK, Jacobson GM et al. The Changes of Target Volumes Through High Dose Rate Brachytherapy for Cervical Cancer When evaluated on High Resolution (3.0) MRI. Practical Radiation Oncology 2012; 2: e101-e106.

23. Kim Y, Modrick JM, Bayouth JE et al. Logistics of tandem and ovoids HDR plan optimization in MRI-guided brachytherapy for cervical cancer: comparisons of standard, graphical, and inverse optimization. Brachytherapy 2009; 8: 121-122.

24. Tanderup K, Pötter R, Lindegaard JC et al. PTV margins should not be used to compensate for uncertainties in 3D image guided intracavitary brachytherapy. Radiother Oncol 2010; 97: 495-500.

25. Tanderup K, Nesvacil N, Pötter R et al. Uncertainties in image guided adaptive cervix cancer brachytherapy: impact on planning and prescription. Radiother Oncol 2013; 107: 1-5.

26. Nesvacil N, Tanderup K, Hellebust TP et al. A multicentre comparison of the dosimetric impact of inter- and intra-fractional anatomical variations in fractionated cervix cancer brachytherapy. Radiother Oncol 2013; 107: 20-25.

27. Morgia M, Cuartero J, Walsh L et al. Tumor and normal tissue dosimetry changes during MR-guided pulsed-dose-rate (PDR) brachytherapy for cervical cancer. Radiother Oncol 2013; 107: 46-51.

28. Anderson C, Lowe G, Wills R et al. Critical structure movement in cervix brachytherapy. Radiother Oncol 2013; 107: 39-45.

29. Jamema SV, Mahantshetty U, Tanderup K et al. Inter-application variation of dose and spatial location of D(2cm(3)) volumes of OARs during MR image based cervix brachytherapy. Radiother Oncol 2013; 107: 58-62.

30. Andersen ES, Noe KO, Sorensen TS et al. Simple DVH parameter addition as compared to deformable registration for bladder dose accumulation in cervix cancer brachytherapy. Radiother Oncol 2013; 107: 52-57.

31. Petric P, Hudej R, Rogelj P et al. Uncertainties of target volume delineation in MRI guided adaptive brachytherapy of cervix cancer: a multi-institutional study. Radiother Oncol 2013; 107: 6-12.

32. Hellebust TP, Tanderup K, Lervag C et al. Dosimetric impact of interobserver variability in MRI-based delineation for cervical cancer brachytherapy. Radiother Oncol 2013; 107: 13-19.

33. Nomden CN, de Leeuw AA, Van Limbergen E et al. Multicentre treatment planning study of MRI-guided brachytherapy for cervical cancer: comparison between tandem-ovoid applicator users. Radiother Oncol 2013; 107: 82-87.

34. Lang S, Nesvacil N, Kirisits C et al. Uncertainty analysis for 3D image-based cervix cancer brachytherapy by repetitive MR imaging: assessment of DVH-variations between two HDR fractions within one applicator insertion and their clinical relevance. Radiother Oncol 2013; 107: 26-31.
Copyright: © 2013 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
 