Impact of interfraction seroma collection on breast brachytherapy dosimetry – a mathematical model

Aashish Bhatt
,
Keith Sowards
,
Geetika Bhatt
,
Andrew Freeman
,
Anthony Dragun

J Contemp Brachyther 2012; 4, 2: 101-105
Online publish date: 2012/06/30
View full text Get citation
 
Purpose: Balloon brachytherapy is a widely accepted modality for delivery of accelerated partial breast irradiation (APBI). Our hypothesis was that inter-fraction seroma collection around the balloon surface would have an adverse effect on dosimetry of the target.

Material and methods: This is a dosimetric re-planning study using two volumetric models (30 cc and 45 cc) in a Contura® multi-lumen balloon (MLB) catheter. In a previously treated patient, two customized baseline plans were generated using multiple channels of the Contura® catheter prescribed to the Planning Target Volume Evaluation (PTV_Eval). Symmetric expansions of 1.0 mm (0-9 mm) increments around the balloon surface were performed to simulate a “Virtual Seroma” (VS) accumulation for both balloon volumes and plans were obtained for each expansion using Eclipse Brachyvision. An analysis of these plans was then performed to evaluate the effect of seroma accumulation on dosimetric parameters of V100 and V90.

Results: 20 plans were generated and analyzed (10 plans for each balloon volume), representing VS of 6.0-66.0 cc. There was a commensurate decrease in the dose delivered to the PTV_Eval V100 and V90 (as defined by the original treatment plan) with increasing VS accumulation leading to a sub-optimal coverage of the PTV_Eval. For 30 cc MLB catheter, V100 decreased by 1.4% and V90 decreased by 0.9% for every 1 cc of VS. For 45cc MLB catheter, V100 decreased by 1.3% and V90 decreased by 1.15% for every 1.0 cc accumulation of VS.

Conclusions: Balloon catheter-tissue adherence ensures daily dose delivery to the planned PTV_Eval. Accumulation of seroma, hematoma or air between HDR fractions can significantly impact PTV_Eval dosimetry. Vacuum-port aspiration prior to delivery of each fraction, if available, should be considered to minimize the risk of geographic under dosing.
keywords:

seroma, dosimetry, breast brachytherapy, contura catheter

 